>>> BrainAmp can be operated on batteries for maximum mobility and mains noise reduction. The automatic standby facility switches BrainAmp MR to energy saving mode (1.5 mA/h) after data acquisition, thereby increasing the battery lifetime considerably. The battery voltage is constantly monitored by the data acquisition program.
>>> Patient safety is guaranteed through possibility for battery operation, opto-coupled signal transfer, and BF classification.
Key Features
Number of channels: 32-256
Input impedance: 10 MOhm
Input noise: 1.5 µVpp
Common mode rejection: 90 dB
Signal range/resolution: +/- 3.2 mV / 100 nV
AD conversion: 16 bit
Bandwidth: 1000 Hz
Sampling rate: 5000 samples/channel/s
Notch filter: 50/60 Hz
Low pass filter: 15 to 1000 Hz
High pass filter: 0.016 Hz / 10 s; up to 16 Hz / 0.01 s
Static discharge: ESD protection
Offset compatibility: +/- 300 mV
Battery running time: Approx. 35 hrs.
Trigger input: 16 bit
Signal transmission: Via twin optical fiber, opto-coupled
Power supply: Integrated power pack
Power input: 130 mA / 1.5 mA standby (32 ch.)
Dimensions (H x W x D): 6 cm x 16 cm x 19 cm (32 channels)
Weight: 1200g (32 channels)
Patient safety: BF, Class I MPG
>>>>
<<<<